Specification:
Assay Method Reaction rate, end point, two-point, double wavelengths, coefficient, multiple standards, blank reagent, blank sample, absorbance at end point, QC, QC graph and turbidity
Temperature System PELTIER temperature control component, 25ºC, 30ºC, 37ºC, accuracy: ±1ºC
Light Source 6V 10W halogen lamp of more than 2000 hours lifespan
GML Precision ±2 nm
FWHM ≤10 nm
Stray Light ≤0.5(absorbance≥2.5)
Absorbance Range -0.214-3.000
Absorbance Linearity ±1%
Absorbance Repeatability ±0.001
Precision ≤1%
Carry Over ≤1%
Resolution 1mAbs/h
Stability ≤3mAbs/h
Flow Cell Permanent quartz flow cell, minimal volume is 32uL, and disposable flow cell is selectable
Sampling System Pressure auto releasing pump ensure the accuracy of suction volume
| Absorption Volume 200uL-3000uL |
| Sample Volume 10uL-50uL |
| Display TFT LCD, touch screen |
| Working condition Temperature 10-30ºC, RH: 20-80% |
| Voltage AC85-264V, auto adapt |
| Power ≤100W |